Variants
- Uro DPT with Utah connector end (Type U)
- Uro DPT with PVB Codan connector end (Type C)
Polycarbonate, Silicon, PVC, Glass, MEMS Sensor
Parameters (MEMS Sensor) | Standard Range |
Operating Pressure in mmHg | 0 to (+)300 mmHg |
Operating Temperature range | 15° to 40°C |
Storage Temperature range | (-)250 to (+)850 C |
Zero Offset | (±)25 mm Hg |
Zero Thermal Drift | (±)0.3 mm Hg/°C |
Calibration requirement | 100 ± 3 |
Sensitivity | 4.95 to 5.0μ V/V/mm Hg |
Excitation Impedance | 330 ohms ± 10% |
Overpressure Tolerance | (-)400 to (+)5000 mm Hg |
Pressure Accuracy | Exceeds AAMI specifications: ANSI/AAMI BP22-1994/(R)2006 |
Operating Range | Non-linearity ± 1.5% of reading |
Nonlinearity and Hysteresis of sensor | -1.5 to 1.5 %VFSS |
Primary pack | Individual Pouch |
Secondary pack | 12 units in an inner carton |
Tertiary pack | Box quantity: 12 x 12 (144 units in a unit shipping box) |