Disposable Diaphragm Domes

Category :
  • Evaluation of the lower urinary tract function through a variety of diagnostic tests including uroflowmetry, cystometry, urethral pressure profile, pressure-flow studies are conducted in patients with lower urinary tract symptoms (LUTS), in conjunction with urodynamic equipment’s like a urine flow meter, urodynamic machine, infusion pumps & pressure measuring equipment.
  • Disposable Diaphragm Domes are used during the above testing procedures. Important components of the pressure dome includes the main dome body, the male & female luer connectors & a membrane diaphragm which snugly fits onto the reusable pressure transducer. This diaphragm acts as an interface between fluid (urine) & the transducer which picks up the pressure wave from the urine flow, which it subsequently converts to an electrical impulse which is displayed as the flow pressures on the monitor.

Disposable Diaphragm Domes

Category :
  • Evaluation of the lower urinary tract function through a variety of diagnostic tests including uroflowmetry, cystometry, urethral pressure profile, pressure-flow studies are conducted in patients with lower urinary tract symptoms (LUTS), in conjunction with urodynamic equipment’s like a urine flow meter, urodynamic machine, infusion pumps & pressure measuring equipment.
  • Disposable Diaphragm Domes are used during the above testing procedures. Important components of the pressure dome includes the main dome body, the male & female luer connectors & a membrane diaphragm which snugly fits onto the reusable pressure transducer. This diaphragm acts as an interface between fluid (urine) & the transducer which picks up the pressure wave from the urine flow, which it subsequently converts to an electrical impulse which is displayed as the flow pressures on the monitor.

Features & Benefits

  • A synchronized pressure transmitting membrane for accurate transmission of pressure waves.
  • Clear transparent body for easy detection of air bubbles & trouble-free quick priming.

Material Specifications

Polycarbonate, High Density Polyethylene, Low Density Polyethylene, Polyurethane, Silicon

Technical Specifications

Parameters Standard Range
Operating Pressure in mmHg 0 to (+)300 mmHg
Operating Temperature range 15° to 40°C
Storage Temperature range (-)250 to (+)850 C
Zero Offset (±)25 mm Hg
Zero Thermal Drift (±)0.3 mm Hg/°C
Calibration requirement 100 ± 3
Pressure Accuracy Exceeds AAMI specifications: ANSI/AAMI BP22-1994/(R)2006

Packing Details

Primary pack Gas permeable membrane sheet & see through laminated film pouch
Secondary pack 50 units in the inner carton
Tertiary pack Box quantity: 200 units (4 cartons)