DISPOSABLE PRESSURE TRANSDUCER KIT

Disposable Pressure Transducer Kit  

Disposable Pressure Transducer Kit

BRAND NAME ACCU +
PRODUCT CODE GP/HM/DPT/9XXXX
Application Surgical procedures (Cardiac/Neurosurgery/critical general surgeries)
  Zoom     Enquiry
 
DESCRIPTION
  • A Disposable Pressure Transducer (DPT) is indicated for pressure sensing during clinical invasive procedures without interfering with the patient’s blood flow and /or blood pressure.
  • The sensor is in fluid communication with different body fluids.
  • By mechanical connection the pressure is transferred to the sensing unit, which translates it to an electrical signal.
  • The stopcock is used for zeroing the pressure monitor before operation.
  • The flush device allows a continuous flow which is required for flushing blood from the monitoring line when connected to the patient assembly.
  • Types of DPT Kits: Single DPT Kits , Double DPT Kit
  • Complete DPT Kit consists of the Following : Disposable Pressure Transducer with inbuilt Flush,Pressure Monitoring Lines, Stopcock,Infusion Set,Dead ender Caps
 
FEATURES & BENEFITS
FEATURES BENEFITS
Use of MEMS sensor chips Accurate transmission of pressure waves
Transparent Housing For easy detection of air bubbles and quick priming
A straight fluid path across sensor Provides easy priming with minimal waveform distortion
Automatic continuous flush/flow@3cc/hr or 30cc/hr Facilitates dynamic response testing of the monitoring system allowing rapid flushing of the Pressure Line after collection of blood sample
Preassembled customized kits User friendly providing easy connection to the patient
Multiple cable end options Compatible with most of the commonly used patient
 
MATERIAL SPECIFICATIONS
COMPONENT MATERIAL
Body Housing & end adaptors Polycarbonate
Release valve system Silicone
Capillary flow system Glass
Luer caps    Polyethylene Polyethylene
 
TECHNICAL SPECIFICATIONS
General Parameters Value/Units Other Units
Pressure Range -30 to 300mmHg -0.58psito 5.8psi (-0.03bar to 0.03 bar)
Overpressure 125psi (X bar) minimum
Electrical Specification Input Excitation 1-10 VDC Calibrated for 6 VDC
Dielectric Breakdown 10,000VDC 5
Risk  Current 2µA Maximum (per AAMI) 5
Input Impedence 1800-3300 Ω     
Output Impedence 285-315 Ω  

Temperature
Compensated 59°F  to 104°F 15°C to 40°C
Operating 59°F to 104°F 15°C to 40°C
Storage -13°to 158°F -25°to70°C
Environmental Humidity 10-90%  
Light Sensitivity 1 mm Hg Maximum per AAMI BP22
Operating Product Life 168 Hours  
Shelf Life   3 Years  
Mechanical Weight <0.0044lb <2g
Volume Displacement <0.0008 <0.02mm3
Media Interface Medical Grade ,Dielectric Gel
Gel Tube Interface Material Polycarbonate
 
PACKING DETAILS
ITEM VALUE
Inner  Cart 4Units
Outer Box 20 units
Box size 610mm x 510mm x 210mm